Characterization of smoke and aerosol deliveries from combustible cigarettes, heated tobacco products and electronic nicotine delivery systems in the Vitrocell® Mammalian 6/48 exposure module

November 3, 2022

https://doi.org/10.1016/j.toxrep.2022.11.001

Brian M. Keyser1, Robert Leverette1, Michael Hollings2, Adam Seymour2, Randy A. Weidman3, Carlton J. Bequette3, Kristen Jordan1
1 RAI Services Company; Scientific & Regulatory Affairs, 401 North Main Street, Winston-Salem, NC 27101, USA
2 Labcorp Early Development Laboratories Ltd., Harrogate, North Yorkshire, UK
3 RJ Reynolds Tobacco Company; 950 Reynolds Blvd., Winston-Salem, NC 27106, USA

 

Highlights

  • Characterization of two 48 well exposure modules with three different product types
  • Comparison of these exposure modules using four different dosimetry techniques
  • Ability to differentiate the aerosol deliveries between different dilution airflows

Abstract
The rapid development associated with Next Generation Tobacco Products (NGTP) has necessitated the development of high throughput methodologies to test their genotoxic potential in vitro when compared to conventional cigarette smoke (CS). An assessment of two Vitrocell® Mammalian 6/48 exposure modules in three independent experiments was made by comparing results from multiple dosimetric techniques applied to aerosol generated from 3R4F Kentucky Reference cigarettes, commercially available electronically heated tobacco product (eHTP) and Electronic Nicotine Delivery System (ENDS) using the Vitrocell® VC10®. Real-time aerosol particle concentration was assessed by means of light scattering photometers and expressed as area under the curve (∑AUC). Nicotine concentrations were determined analytically by LC/MS. Humectant amount and distribution was assessed for eHTP and ENDS by the quantification of free glycerol in a phosphate buffered saline (PBS) trap, whereas total particulate matter (TPM) was assessed in the 3R4F cigarettes by the fluorescence of the particulate at 485nm in anhydrous dimethyl sulfoxide (DMSO) trap within the exposure.
Dose was adjusted by means of the addition of ambient air to dilute the whole smoke/aerosol in L/min and sampled into the system at a rate of 5 mL/min. Dilution of CS ranged from 8.0 to 0.5 L/min and for the eHTP and ENDS ranged from 4 to 0 L/min (undiluted). Dosimetric analysis of the system showed good concordance within replicates (p-values ranged from p=0.3762 to p=0.8926) and showed that the Vitrocell® Mammalian 6/48 is a viable means for genotoxic assessment of aerosol generated from both conventional cigarettes and NGTP. Results demonstrate the need to tailor dosimetry approaches to different aerosols due to variations in the physio-chemical composition, with a multi-dosimetry approach recommended.